1 Summer
|
1.1 Equipment name
|
Xenon lamp aging tester
|
1.2 Equipment model
|
BND-HXD-800L
|
1.4 Specimen limitation
|
Do not use explosive substances, combustible substances, or materials containing such substances as specimens. Otherwise, there is a risk of explosion, fire, or other accidents.
●Do not allow corrosive substances to enter the tank. Use of corrosive substances or humidified water, especially stainless steel, resin, silicon, etc., may cause a significant decrease in the corrosion resistance of the product.
●Do not use living organisms or substances that will exceed the permissible heat generation as specimens.
|
1.5 Equipment use
|
The xenon lamp aging tester uses a xenon arc lamp that mimics the full sunlight spectrum to reproduce the destructive light waves that exist in different environments, and can provide the appropriate environmental simulation and accelerated testing for scientific research, product development and quality control. It can be used for the selection of new materials, the improvement of existing materials or the evaluation of changes in durability after changes in the composition of materials, can be a good simulation of the changes in different environmental conditions, materials exposed to sunlight.
|
1.7 Standard
|
GB/T16422.2 Plastics Laboratory light exposure test method part 340, ASTM G155 part and other standards!
|
2 size specification
|
2.1 Inner box size
|
W800*H800*D800mm (drum diameter 500mm sample fixture width 75*height 150*thickness 5mm can put about 40)
|
2.2 Outer case size
|
W1350*H2050*D1850mm subject to the object.
|
2.3 Sample condition
|
Sample placement test space is not more than three quarters of the total space, no heat load.
|
3 Equipment Performance
|
3.1 Temperature range inside the box
|
0℃~80℃
|
3.2 Blackboard temperature range
|
+45℃~100℃(±3℃)Depending on the temperature inside the box and the irradiation intensity settings
|
3.3 Box temperature deviation
|
±2℃(without light on)
|
3.4 Temperature resolution
|
0.1°C
|
3.5 Temperature Fluctuation
|
±0.5℃(without light)
|
3.6 Relative humidity range
|
20% to 95% R.H (Controlled range chart in purple area below)
|
3.7 Relative humidity deviation
|
±3%R.H (without light)
|
3.8 Relative humidity fluctuation
|
±2.5%R.H (without light)
|
3.9 Light source type
|
Domestic 1 water-cooled xenon lamp wavelength 280-800nm, power 4.5KW, average service life of 1200 hours, equipped with a set of light monitoring and feedback sensors, does not contain a filter. (The equipment comes with lamp cooling circulating water)
|
3.10 Irradiation intensity
|
@340nm irradiation intensity 0.35 ~ 0.6W/m2 adjustable
|
3.11Rotation speed
|
1-9RPM
|
3.12 Rain function
|
Support internal and external spray, can be indirect spray, continuous spray.
|
3.13 Function Mode
|
Irradiation mode, dark mode, spray mode can be alternately cycle operation.
|
※The above performance was measured under the conditions of ambient temperature 25℃±3℃, relative humidity ≤85% unloaded without specimen.
|
4 structural characteristic
|
4.1 Controller characteristics
|
(1) Hardware architecture: PLC listed Xindjie XD5 series + MCGS 7-inch human-machine interface touch screen
(2) Software developed independently by BONAD (expandable and powerful, can be customised according to the requirements of non-standard change software)
(3) Flat design, easy to operate.
|
4.2 Control mode
|
●Programmes: 50 groups of programmes, each group of programmes has 50 segments, you can set all cycles of the programme, part of the cycle, programme connection, set the end-of-programme action, set the standby action.
●Determined value: running time and running conditions can be set.
|
4.3 Sensors
|
PT100 3-wire high accuracy sensor
|
4.4 Analogue output
|
Voltage pulse (SSR) 2 points, DC 24V, minimum pulse width 5ms.
|
4.5 Communication interface
|
●USB: The controller can download historical curves and historical data, historical data is saved in Excel format, you can download the required data by date. Historical data is saved in JPG image format, which can be enlarged, reduced and then downloaded as required.
●RS485 interface
|
5 safety device
|
5.1 Electrical control system
|
●Heating and humidification over-temperature protection ●Overload short circuit leakage protection ●Blower over-current protection ●Lamp over-temperature protection
●Water shortage protection ●Power shortage reverse phase protection ●Shower pump overcurrent protection
|
5.2 Refrigeration system
|
●Compressor overcurrent protection ●Compressor overpressure protection ●Compressor overtemperature protection
|
6 Environmental conditions of use
|
6.1 Equipment placement
|
No strong vibration around the equipment
No strong electromagnetic field influence around the equipment
No flammable, explosive, corrosive substances and dust around the equipment.
Appropriate space for use and maintenance around the equipment, as shown in the figure below:
A: not less than 10cm
B: not less than 60cm
C:Not less than 60cm
D: not less than 150cm
|
6.2 Environmental conditions
|
Temperature: +5℃~30℃ Relative humidity: ≤85% Barometric pressure: 86kPa~106kPa
|
6.3 Power supply
|
Voltage: AC380V±10%, frequency: 50HZ, protective earth ground resistance less than 4Ω.
Users in the installation site for the equipment configuration of the corresponding capacity of the air or power switch, and this switch must be independent for the use of this equipment (not connected to the socket use)
Installed power: 18KW Maximum current: 27A
|
6.4 Water
|
Pure or distilled water, not unfiltered tap water;
|